Ac Voltage; Crest Factor (Cf) - Chauvin Arnoux F23 Manual De Instrucciones

Pinza de potencias
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

www.motralec.com / service-commercial@motralec.com / 01.39.97.65.10
PEAK Mode
An additional error of 0.5 % on peak value, is to be added to the error specified in RMS. Residual current on peak
value: 0.3 A max.
RECORD PEAK Mode
An additional error of 1 % on peak value, is to be added to the error specified in RMS. Residual current on PEAK
value : 0.5 A max.

5.3 AC VOLTAGE

Particular reference conditions : sinusoidal signal
!
Limits on voltage measurements
- Operating frequency : 10 Hz to 5 kHz
- Permitted permanent overload : 1.5 kV PEAK
!
NB : Display of
Measurement range
V AC
Accuracy
Resolution
Input impedance
May be used up to 1500 V PEAK on condition that the corresponding rms value is < 600 V (IEC 1010).
(1)
PEAK Mode
An additional error of 0.5 % on peak value, is to be added to the error specified in RMS. Residual voltage on peak
value: 0.3 V max.
RECORD PEAK Mode
An additional error of 1 % on peak value, is to be added to the error specified in RMS. Residual voltage on PEAK
value: 0.3 V max.

5.4 CREST FACTOR (CF)

Particular reference conditions
- Max peak value 1500 V or 1500 A
- Min value of signals 5 V and 5 A
- Operating frequency 40 Hz to 70 Hz
Measurement range
CF
Accuracy
Resolution
: - steady above 600 V RMS
- flashing above 1500 V PEAK
0.05...59.99 V
1% rdg ± 10 counts
10 mV
1.00...3.50
2% rdg ± 2 counts
60.0...599.9 V
1% rdg ± 2 counts
100 mV
1MΩ
3.51...5.99
5% rdg ± 2 counts
0.01
27
600 V RMS
or 600...1500 V peak
(1)
1 V
6.00...10.00
10% rdg ± 2 counts

Publicidad