Daily Disinfecting - Omron CompAir Elite NE-C30 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

You can disinfect daily by soaking the parts in a vinegar solution or using a
commercially available medical disinfectant. If your physician or respiratory
therapist specifies a different cleaning procedure follow their instructions.
Disinfect the nebulizer kit and the mask, nosepiece or mouthpiece
after the last treatment of the day.
VINEGAR METHOD
1. Make a solution of 1 part white vinegar and 3 parts distilled water.
Make enough solution to submerge the parts.
2. Remove the inhalation accessory (mouthpiece, mask, or optional
nosepiece) from the nebulizer kit.
3. Disconnect the air tube from the nebulizer.
4. Gently twist the inhalation top counterclockwise and lift to separate
the nebulizer into two sections.
5. Remove the baffle.
6. Wash all the parts of the nebulizer kit, the mouthpiece, optional nosepiece
or mask in warm water and a mild detergent. Rinse in hot tap water.
You may also wash the nebulizer kit, mouthpiece or nosepiece in a
dishwasher. DO NOT wash the mask in a dishwasher.
CAUTION
Place the nebulizer pieces in a basket recommended for
dishwasher use. Place the basket on the top shelf to avoid
any damage to the nebulizer kit.
7. Submerge the parts in the white vinegar and distilled water solution
for 30 minutes.
8. Remove the parts and discard the solution. Rinse the parts with hot
tap water.
9. Hand dry or air dry in a clean environment using a soft, clean,
lint-free cloth.

DAILY DISINFECTING

18