Technical Specifications; Device Disposal - FLAEM Hospi Neb JUNIOR F2000 Manual De Instrucciones

Ocultar thumbs Ver también para Hospi Neb JUNIOR F2000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38

TECHNICAL SPECIFICATIONS

Mod. FLAEM JUNIOR F2000
Voltage:
230V~ 50Hz 210VA
Max Pressure:
Compressor air output:
Noise level (at 1 m):
Operation:
Dimensions
Weight
Operating conditions:
Storage conditions:
Operating/storage
atmospheric pressure:
APPLIED PARTS
Type BF applied parts include:
RF6 Plus NEBULIZER
Medication minimum capacity:
Medication maximum capacity: 8 ml
OPERATING
(2)
PRESSURE bar
1.2
Data shown is in accordance with Flaem Nuova internal procedure I29-P07.5.
(1)
Research GmbH Co. KG on behalf of TÜV Rheinland LGA Products GmbH – Germany in compliance with the new European Standard
for aerosol therapy units, Standard EN 13544-1. Further details are available upon request.

DEVICE DISPOSAL

In conformity with Directive 2012/19/EC, the symbol shown on the appliance to be
disposed of indicates that this is considered as waste and must therefore undergo
"sorted waste collection". The user must therefore take (or have taken) the above
waste to a pre-sorted waste collection centre set up by the local authorities, or else
give it back to the dealer when purchasing a new appliance of the same type. Pre-sorted
waste collection and the subsequent treatment, recovery and disposal operations favour
the production of appliances made of recycled materials and limit the negative effects
of any incorrect waste management on the environment and public health. The unlawful
disposal of the product by the user could result in administrative fines as provided by
the laws transposing Directive 2012/19/EC of the European member state in which the
product is disposed of.
115V~ 60Hz 230VA
3.5 ± 0.5 bar
14 LPM approx
57 dB (A) approx
Continuous use
20(W)x18(D)x11(H) cm
1.8 Kg
Temperature: min 10°C; max 40°C
Air humidity: min 10%; max 95%
Temperature: min -25°C; max 70°C
Air humidity: min 10%; max 95%
min. 69KPa; máx. 106KPa
(C2,C3,C4,C5)
2 ml
MMAD
BREATHABLE FRACTION
(2)
μm
< 5 μm (FPF) %
2.8
76
220V~ 60Hz 230VA
DELIVERY ml/min
(1)
open valve-
holder MAX
(approx.)
0.54
In vitro characterization carried out at the Inamed
(2)
14
100V~ 50/60Hz
230VA
closed valve-
holder MIN
(approx.)
0.31

Publicidad

Tabla de contenido
loading

Tabla de contenido