D. ABSENCE OF AIR DELIVERY
Message
Reason
No air delivery
ACP tubing set leaking or not well
while green LED
connected.
is ON.
Output connector
malfunctioning.
Internal components
malfunctioning.
E. LEAKAGE AFTER CORNEAL TISSUE SETTING ON THE ARTIFICIAL CHAMBER
Message
Reason
Leakage when
ACP tubing set not properly
the corneal
connected or artificial chamber
tissue is on
leaking.
the artificial
Output valve malfunctioning.
chamber
Internal components
malfunctioning.
XII. CARE & MAINTENANCE
A. CARE
To clean ACP system, wipe the system with a lint-free cloth moistened with disinfecting solutions. Do not use
solvents, abrasives, fluids or solutions on ACP system, ACP footswitch and ACP power supply.
Avoid any liquid projection on ACP system, ACP footswitch and ACP power supply.
Do not open neither ACP system nor its accessories.
In cases of emergency, disconnect the power socket or the power supply connector.
In cases of unusual vibrations or noises, instability, unusual performance, do not use ACP system and contact
MORIA or your distributor for fixing.
In order to assess any variation of the system accuracy, it is recommended to check on a routine basis the
obtained pressure inside the artificial chamber with a 65 and/or 90 mmHg Barraquer tonometer and ACP
system set up at 65 and/or 90 mmHg accordingly.
MORIA strongly recommends having the system thoroughly inspected by MORIA on a routine basis every year.
Please contact MORIA for any other servicing.
B.MAINTENANCE

1. REPLACING FUSE

Steps
What to do
1
Switch OFF ACP system
2
Unscrew the fuse on ACP rear panel
3
Gently pull the fuse on ACP rear
panel
4
Remove the cartridge
5
Change the fuse
6
Gently push the fuse
7
Screw the fuse on ACP rear panel
Intervention
Disconnect ACP tubing set and place your finger on the output connector
to check whether air is delivered outside or not:
• if yes: then change ACP tubing set.
• if unable to solve the problem: do not use ACP system and contact
your distributor or MORIA to have it checked.
Intervention
Check if the corneal tissue has not been damaged after collapsing.
Change tubing and start again the different steps described in
paragraph VIII.
IMPORTANT WARNINGS
T1.6AL250V (5x20mm)
#65112-D-12.2017
MORIA
Related
pictures
E1
E2
E3
E4
E5
E6