2. "CLOSED SYSTEM" = PRESSURE PREVIOUSLY ESTABLISHED BY ACP SYSTEM

Steps
What to do
13-2a
Before proceeding to
the operation, check the
pressure display.
13.2b
Establish pressure in
the artificial chamber
by shifting the stopcock
of ACP tubing set to be
open to both ACP system
and artificial chamber,
while being closed to the
syringe.
13-2c
Close the system by
shifting the stopcock of
ACP tubing set to be open
to both ACP system and
syringe, while being closed
to the artificial chamber.
13-2d
Corneal tissue is ready for
operation.
D. REMOVE THE CORNEAL TISSUE AFTER OPERATION

1. "OPEN SYSTEM"

Steps
What to do
14-1a
Shift stopcock of ACP
tubing set to be open to
both artificial chamber and
syringe, while being closed
to ACP system.
14-1b
Push on the footswitch to
close the output valve.
14-1c
Remove the corneal tissue.

2. "CLOSED SYSTEM"

Steps
What to do
14-2a
Proceed to the operation
by shifting the stopcock of
ACP tubing set to be open
to both artificial chamber
and syringe, while being
closed to ACP system.
14-2b
Push on the footswitch to
close the output valve.
14-2c
Remove the corneal tissue.
IMPORTANT WARNINGS
In case pressure display is different from desired value, readjust the
pressure value on the digital display by turning the knob.
Visually check that the displayed value is similar to the desired one. If not,
readjust the pressure value on the digital display by turning the knob.
Note: ACP system can deviate during a few minutes of ± 15 mmHg. After
warm-up time, ACP system becomes stable.
Note: once the system is closed, syringe can gently be pushed away.
IMPORTANT WARNINGS
Visually check that:
• the yellow LED is ON and fixed.
• the green LED is OFF.
IMPORTANT WARNINGS
Visually check that:
• the yellow LED is ON and fixed.
• the green LED is OFF.