OHAUS VE Serie Guia Del Usuario página 3

Ocultar thumbs Ver también para VE Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 5
EN-2
2.2 Connecting an Indicator
Certain Floor Scale systems will have an Indicator already attached to the cable end. If the Indicator is not attached or if the
Indicator is purchased separately follow the steps listed below to make the proper connections.
FLOOR SCALE WIRING COLOR CODE
Indicator Signal Name
EXC+ (IN+)
SEN+
SIG+ (OUT+)
SIG- (OUT-)
SEN-
EXC- (IN-)
Shield (GND)
1.
Route the Floor Scale cable to the position where the Indicator will be mounted.
2.
Consult the installation instructions of the Indicator to properly attach the cable connections inside the
Indicator,
3.
The Signal names of the Floor Scale wires are shown in the table above, Floor Scale Wiring Color Code.
CAUTION: Ensure that the cable is protected against possible damage when routed to the Indicator.
3. OPERATION
Operating Limits
MAXIMUM ADMISSIBLE LOADING.
Model
VE1500S/R/L
VE3000R/L
VE1500SW/RW/LW
VE3000RW/LW
For continued best operating performance:
Avoid falling loads, shocks and hard lateral impacts.
When weighing, all items should be centrally located on the floor scale and
not over hanging the sides or the ramps.
Verify the calibration of the floor scale system at regular intervals.
Maximum Central
1
load
(80% Surface Area)
2500 kg
3800 kg
2500 kg
3800 kg
U-Pallet Scale Wire Color
Green
Yellow
White
Red
Blue
Black
Yellow/Green
2
Maximum Side load
(30% Surface Area)
1500 kg
1800 kg
1500 kg
1800 kg
VE Series Floor Scale
One Side Corner
3
Load
(10% Surface Area)
1000 kg
1200 kg
1000 kg
1200 kg

Publicidad

Tabla de contenido
loading

Tabla de contenido