Precision & Analytical Series
Steps for linearity calibration
1. Press and hold the
will display.
2. Press Yes to enter the Calibration Menu.
3. To change the calibration mode, press No until LineRr
(linearity calibration) is displayed.
4. The calibration mass value will be shown in the display. After
the display shows 100.0000 g, please place weight(s) of 100
g on the pan for calibration.
5. Remove the weight of 100 g from the pan. After a while,
200.0000 g will be displayed on the screen. Please place
weight(s) of 200 g on the pan.
6. Once the linearity calibration is completed successfully,
CAL.Done will display.
Press any button to return to the previous screen.
7. Remove the weight, and the reading will be set to zero.
button, and the Calibration Menu
17