RF Proximity
385 Pulse Modulation, 18 Hz
(IEC 61000-4-3)
450 FM + 5Hz deviation: 1 kHz sine
710
745
780 Pulse Modulation: 217 Hz
810
870
930 Pulse Modulation: 18 Hz
1720
1845
1970 Pulse Modulation: 217 Hz
2450 Pulse Modulation: 217 Hz
5240
5500
5785 Pulse Modulation: 217 Hz
Surges on AC
Common Mode: ± 2 kV
mains lines
Differential Mode: ± 1 kV
(IEC 61000-4-5)
Power frequency
30 A/m
magnetic field
50/60 Hz
(IEC 61000-4-8)
Voltage dips
Dip > 95%,
and short
Dip > 95%,
interruptions on
Dip 30%,
AC mains input
lines
Drops >95%, 250 cycles (50 Hz)
(IEC 61000-4-11)
0.5 cycles
1 cycle
25 cycles (50 Hz)
30 cycles (60 Hz)
300 cycles (60 Hz)
27 V/m
Equipment with high RF emissions
28 V/m
should be kept at a distance to reduce
the likelihood of interference.
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
± 2 kV
Mains power quality should be that
± 1 kV
of a typical commercial or hospital
environment.
30 A/m
Equipment that emits high levels of
power line magnetic fields (in excess of
3A/m) should be kept at a distance to
reduce the likelihood of interference.
0.5 cycles
Mains power should be that of a typical
1 cycle
commercial or hospital environment. If
25 cycles (50 Hz)
you require continued operation during
30 cycles (60 Hz)
power mains interruptions, ensure that
250 cycles (50 Hz)
batteries are installed and charged.
300 cycles (60 Hz)
Ensure that battery life exceeds
longest anticipated power outages
or provide additional uninterruptible
power source.