Skillguide

Consistent and objective feedback on performance during
practice helps develop and reinforce skills. The Skillguide
provides feedback to the instructor and student to support the
student's initial skill development by means of a colour light
display.
Inflation
When lung inflation volume reaches 0.7 l (standard ventilation)
or 0.4 l (O2 enriched ventilation), the green signal light will be
activated. If lung inflation volume exceeds 1.0 l (standard ven-
tilation) or 0.6 l (O2 enriched ventilation), an orange warning
light is activated. A red warning light indicates a too fast infla-
tion.
To get correct feedback when performing standard ventilation,
the larger lung (15 22 50) has to be used.
To get correct feedback when performing O2 enriched ventila-
tion, the smaller lung (15 26 00) has to be used.
Chest compression
When the breastbone area of the chest is depressed about
4 cm (at least 1 1/2"), the green signal light is activated.
If compression exceeds 5 cm (2" ) an orange warning light
is activated.
Incorrect hand position
Sensors are located under the switch cover mounted on the
underside of the chest cover. The red warning light will be
activated if the chest is depressed with hands placed outside
the correct area.
Light
Inflation volume:
green
orange
Compression depth:
green
orange
Wrong hand position:
red
Too fast inflation:
red
*Measurement tolerance in the manikin: +/- 15% in normal room temperature.
Measurements based on ILCOR guidelines. O2 enriched ventilation requires airway 15 26 00 for correct
feedback.
To use
See 2) under Unpacking and set up on page 6.
The Skillguide will turn "off" automatically if no activity takes
place for 5 minutes.
Note: If the Skillguide is turned "on" and the red signal light
for wrong hand position comes on, the Skillguide is not com-
pletely plugged into the manikin.
Power Supply
The four 1.5 V alkaline batteries, size C, will provide approxi-
mately 125 hours of use in the classroom. This will vary
depending on use patterns and brand of batteries.
Connection Cable
This is permanently connected to the Skillguide and has an
8 pin male connector that is to be connected to the manikin.
The connector has been moulded for proper orientation prior
to plugging it into the manikin's side.
Caution
Make sure that the Skillguide
is unplugged from manikin:
- when airway is changed, or
- when internal service is carried out.
Indication
Limits*
correct
0.7 l - 1.0 l
0.4 l - 0.6 l
too much
>1.0 l
>0.6 l
correct
4 cm (1 1/2") - 5 cm (2")
too much
>5 cm (2")
Pressure outside
of acceptable area
Flow > 1 l/sec.
Flow > 0.5 l/sec.
7
(standard)
(O2 enriched)
(standard)
(O2 enriched)
(standard)
(O2 enriched)
Resusci Anne Modular System