A typical test report will look as:
MOISTURE TEST
FILE NAME: RES00032
DATE:
13/12/2008
TIME:
09:38:07
SER NO:
AE12345678
TEST NO:
_ _ _ _ _ _ _ _ _
USER NO:
9876543210
Preset:
PST 05
Bread-4
Result:
% Moisture
Heating:
Step
Temp:
Step 1 120 C
Time 1 04:00
Step 2 110 C
Time 2 90:00
Rapid
On
Interval:
30 Sec
Stop:
Time/Stable
60 Min
0.002 g
30 Sec
Start:
Manual
INIT MASS:
12.341 g
MODE TEMP
TIME
Step0
65C
00:30
Step0
159C
01:00
Step0
157C
01:30
Step0
158C
02:00
Step0
156C
02:30
Step1
157C
03:00
Step1
149C
03:30
Step1
140C
04:00
Step2
131C
04:30
Step2
120C
05:00
Step2
111C
05:30
*********** AUTO STOP ***********
LAST TEMP:
111C
TEST TIME:
05:30 Min
FINAL MASS:
11.820 g
MASS LOSS:
0.521 g
RESULT:
4.27 %M
English language version shown in the example.
 
58
|
P a g e
Heading
File name used by analyzer to store result
From Analyzer
Serial number
A blank line for users to fill in a number if they wish.
User number
Preset method if used.
Test settings
RESULT
1.26 %M
2.11 %M
3.15 %M
3.79 %M
Measured initial mass.
4.11 %M
4.19 %M
Test Heading
4.22 %M
Results printed once every 30 seconds as set by the
4.24 %M
INCR time. Note temperature during first part of test is
4.26 %M
120C X 130% as the Rapid function was enabled.
4.27 %M
4.27 %M
Step0 signifies the rapid function is active. After 3
minutes the Step1 value is the target.
TIME STOP if timed out or MAN STOP if user stopped
Summary of the test
© Adam Equipment Company 2017