IB-BPA3000-0912-01:Layout 1 09/01/2013 17:43 Page 15
Table 2 Guidance and manufacture's declaration – electromagnetic immunity – for all
ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration – electromagnetic immunity
The BPA-3000-EU is intended for use in the electromagnetic environment specified below. The customer of
the user of the BPA-3000-EU should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC
61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
Power frequency
(50Hz) magnetic
field IEC
61000-4-8
NOTE: U T is the a.c. mains voltage prior to application of the test level.
IEC 60601 test
Compliance level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2kV for power
supply lines
supply lines
±1 kV line(s) to
±1 kV differential
line(s)
mode
<5% U T
<5% U T
(>95% dip in U T )
(>95% dip in U T )
for 0.5 cycle
for 0.5 cycle
40% U T
40% U T )
(60% dip in U T )
(60% dip in U T )
for 5 cycles
for 5 cycles
70% U T
70% U T
(30% dip in U T )
(30% dip in U T )
for 25 cycles
for 25 cycles
<5% U T
<5% U T
(>95% dip in U T )
(>95% dip in U T )
for 5 sec
for 5 sec
3A/m
3A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floor are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the BPA-3000-EU requires
continued operation during power mains
interruptions, it is recommended that the
BPA-3000-EU be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
15