Troubleshooting; Specifications - VWR CO310 Manual De Operaciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 23

TROUBLESHOOTING

MAIN DISPLAY
OvEr
OvEr/Undr during calibration
MAIN
SECONDARY
DISPLAY
DISPLAY
OvEr/Undr

SPECIFICATIONS

Display
Conductivity,
0.0 to 499.9 uS/cm
Auto-ranging
500 to 4999 uS/cm
5.00 to 49.99 mS/cm
50.0 to 200.0 mS/cm
Salinity
Temperature °C
Reference Temperature
Temperature Coefficient
Cell Constant
TDS Constant Range
Power/Battery Life
Calibration Back-up
Audio Feedback
Instrument Case
Operating Temp. Range
Operating Relative Humidity Range
Temperature Probe
Dimensions (L x W x H)
Weight (batteries included)
• Conductivity is >200.0 mS
• Salinity is > 70.00 ppt
Cell Constant Calibration is
OvEr
Temperature >90.0 °C
Undr
Temperature < -10.0 °C
Range
0.0 to 70.0 ppt
-10.0 to 90 °C
15.0 to 25.0
0.0% to 4.0%
5.00
0.30 to 1.00
One 9V battery/ Approximately 500 hours
Yes
Yes, on all touch keys
Waterproof when connector cap installed, IP 67
0 to 50
up to 95%
Thermistor, 10kΩ / 25
18.7 cm x 7.6 cm x 3.8 cm (7.37 in x 3 in x 1.5 in)
270 grams (.6 lb)
PROBLEM
out of range
Accuracy
±1% of reading plus 2 uS/cm
±1% of reading plus 5 uS/cm
±1% of reading plus 0.05 uS/cm
±2.5% of reading plus 0.5 mS/cm
0.2% Full Scale
±0.2 °C or ±0.4% Full Scale,
whichever is greater
°C
±
0.50
°C
°C
7
POSSIBLE SOLUTION
• Completely submerge the
probe.
• Allow sufficient time for the
electrode and Temp probe
stabilization.
• Recalibrate with correct value
for the conductivity standard.
• Replace conductivity standard.
• Clean cell.
• Return for service.
• Decrease/Increase the sample
temperature.
• Return for service.
Resolution
0.01 uS/cm
1 mS/cm
0.01 mS/cm
0.1 mS/cm
0.1 ppt
0.1 °C