ResMed S8 Elite Manual De Instrucciones página 39

Ocultar thumbs Ver también para S8 Elite:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Problem
Possible Cause
Display error message:
You have experienced
excessively high leak levels for
HIGH LEAK!!!
more than 20 seconds.
Adjust Mask
Display error message:
Component failure.
Exxxx (where xxxx
defines an error)
The following message
The Data Card is not inserted
is displayed on the
correctly.
LCD after you try to
update settings using
the Data Card:
Card Error
Remove Card
The Data Card is only partly
inserted.
You may have removed the Data
Card before settings were copied
to the S8 Elite.
The card is not a ResScan Data
Card.
The following message
The Data Card is not inserted
is displayed on the
correctly.
LCD after you try to
copy data to a Data
The card is not a ResScan Data
Card:
Card.
Card Error
Remove Card
You may have removed the Data
Card before data was copied to
it.
The following message
The identification details on the
is displayed on the
Data Card do not match the
LCD after you try to
details on your device.
update the settings
using the Data Card:
Settings Invalid
Remove Card
Solution
Check that your air tubing is
connected properly and that your
mask does not leak excessively.
Use the mask-fitting feature to
help you to fit your mask
properly.
If this message appears again,
contact your clinician.
Return your S8 Elite for servicing.
Ensure that the Data Card is
inserted with the arrow facing up
and pointing towards the device.
Ensure that the device is switched
on and in standby mode before
you insert the Data Card.
Ensure that the Data Card is
inserted into the slot on the Data
Card module as far as it will go.
Reinsert the Data Card and wait
for the Settings Success
Remove Card message to
appear on the LCD.
Only use a ResScan Data Card.
Ensure that the Data Card is
inserted with the arrow facing up
and pointing toward the device.
Use a ResScan Data Card, as
specified in this manual.
Reinsert the Data Card and wait
for the Copy Complete
Remove Card message to
appear on the LCD.
Contact your clinician
immediately.
T
ROUBLESHOOTING
35

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido