TRAVEL INFORMATION
NOTE – For details on part numbers for alternate power sources, refer to Replacement Items/Spare Parts.
Air Travel
Your device is RTCA/D0-160 Section 21 Category M approved for use on commercial airlines when using battery power. You must turn off the Bluetooth. Refer to Features & Settings -
Add-Ons in this guide for details on Bluetooth.
International Travel
Your device is capable of accepting line voltages of 100-240V~, 50/60Hz. For travel to another country, simply order the correct power cord for that region from your provider or use
proper adapter.
Altitude
This device automatically compensates for altitude between 1060 hPa and 700 hPa (~ 1,400 ft below sea level to 9,800 ft). At higher altitudes, breathing patterns change and you may
lose some benefit of your CPAP therapy.
Battery/DC Operation
Several DC power source options are available:
1.
Your device can be operated from a 12 V DC power source by using the optional DC Cable.
2.
Lastly, instead of connecting12V DC power directly to the CPAP device, you can pass the DC power through an AC Inverter that will deliver AC power to the system. The power
rating of the inverter must be at least 200 watts @ 110V~ / 400 watts @ 220V~.
Approximate Battery Operation Run Time
Example Battery Size = 100 W/Hr
CPAP Setting (cmH20)
Minimum Run Time (Hrs)
5
10
15
20
NOTE – Run times are at a breath rate of 20 breaths per minute and standard mask leak.
A-DV64
CPAP Only
Heater Setting = 3
Minimum Run Time (Hrs)
16.9
8.7
11.9
7.5
8.7
6.4
6.8
5.8
CPAP with Heated Humidifier
Heater Setting = 5
Minimum Run Time (Hrs)
5.6
4.7
4.4
4.0
CPAP with Heated Humidifier and PulseDose Module
Heater Setting = 3
Heater Setting = 5
Minimum Run Time (Hrs)
Minimum Run Time (Hrs)
9.5
7.8
7.1
6.1
5.9
5.4
4.7
4.4
EN - 19