The flow setting of the oxygen source must be specified by a physician.
Maximum oxygen pressure is 50 psi. Maximum oxygen flow rate is 10 lpm.
Always turn on CPAP before turning on oxygen flow.
Always turn off oxygen flow before turning off CPAP.
Proper filter function is important for the operation of the device and to protect the device from
damage.

INTRODUCTION

Intended Use

The DeVilbiss Intellipap2/DeVilbiss BLUE Series is intended for use in treating OSA in spontaneously
breathing patients 30 Kg (66 lbs) and above by means of application of positive air pressure. The
Device is to be used in Home and Healthcare Environments.
Contraindications
Positive airway pressure therapy may be contraindicated in some patients with the following pre-
existing conditions:
severe bullous lung disease
pneumothorax
pathologically low blood pressure
dehydration
cerebrospinal fluid leak, recent cranial surgery or trauma
a bypassed upper airway
hypoventilation
The use of positive airway pressure therapy may be temporarily contraindicated if you exhibit signs of a
sinus or middle ear infection. Contact your physician if you have any questions concerning your therapy.
MR Unsafe
Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause
unacceptable risk to the patient or damage to the CPAP or MR medical devices. The device and
accessories have not been evaluated for safety in an MR environment.
Do not use the device or accessories in an environment with electromagnetic equipment such as CT
scanners, Diathermy, RFID and electromagnetic security systems (metal detectors) as it may cause
unacceptable risk to the patient or damage to the CPAP. Some electromagnetic sources may not be
apparent, if you notice any unexplained changes in the performance of this device, if it is making
unusual or harsh sounds, disconnect the power cord and discontinue use. Contact your home care
provider.
This device is for single patient reuse only.
EN - 6

SET UP

First, unpack your CPAP and identify all items.
Contents
9
8
7
6
1. Carrying Case
2. Tubing
3. Filter Cover
4. Extra Air Inlet Filters
5. Optional Fine Particle Filter
6. SD Card (optional)
7. Power Supply
8. Power Cord
9. CPAP Device
Instruction Guide (not shown)
NOTE - Contents vary by model. Please refer to the contents list on the
device carton for items specific to your model.
NOTE - For models with heated humidification option, please refer to
humidifier guide for device set up.
1
2
3
4
5
A-DV64