Figure 6-14
Figure 6-15
5) Wireless Transmission
When select "Wireless", if the device has built-in wireless module, press the confirm key to achieve turning on/off
the wireless module, when it is ON, it can transmit data.
6) Scaler Operation(Calibration)
Select "Scaler (Calibration)" to its sub-menu as Figure 6-16, then the scaler (Calibration) volume can be selected,
after selected the volume, it will enter scaler (Calibration) interface as Figure 6-17. In this interface, when moving
the scaler once, the interface will display "REPEAT", then move the scaler continuously. After twice right scaler
(Calibration), the scaler (Calibration) is successful, then the interface will display "OK!".
Finally the interface will jump to the former interface before scaling (Calibration). If Figure 6-18 appears, it indicates
the scaler (Calibration) is error, then repeat scaling (Calibration) until succeeding. If Figure 6-19 appears, then
please confirm the volume between scaler (Calibration) and selected is accordant, then scaling (Calibrate) until
succeeding. If you need to stop scaling (Calibration), only press the confirm key to exit to the former interface
before scaling (Calibration). (The interface before scaling (Calibration) refers to: If scaling (Calibrate) after com-
pleting measuring, then it will exit to the control setting interface, otherwise jump to the measurement interface
as Figure 6-1).