Guidance and manufacturer's declaration – electromagnetic immunity – for EQUIPMENT and SYSTEMS
that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The Spirometer is intended for use in the electromagnetic environment specified below. The customer or the user
of Spirometer should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Radiated
3 V/m
RF
80 MHz
IEC61000-4-3
to 2.5 GHz
Compliance
Electromagnetic environment - guidance
level
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the Spirome-
ter, including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
3.5
d=
E
7
d=
E
Where P is the maximum output power rating of the tran-
smitter in watts (W) according to the transmitter manu-
facturer and d is the recommended separation distance
in metres (m).
ENGLISH
√P
80 MHz to 800 MHz
1
√P
800 MHz to 2.5 GHz
1