Baxtran BR16 Manual De Instrucciones página 63

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

SPEED 1
•Use the key
LF 4
•Use the keys
BLINF
LF 5
DISPLAY
ZP 0 OFF
LF 5
ZP 1 One division will not show being at zero
ZP 2 Two division will not show being at zero
ZP 3 Three divisions will not show being at zero
ZP 4 Four divisions will not show being at zero
ZP 5 Five divisions will not show being at zero
*This function is going to be blocked when UF-5 is in mode HOLD 1
*The value of factory is ZP 0
DISPLAY
LF 5
• Press the key
automatically.
ZP 0
• Use the key
ESC to exit from the configuration and go back to the menu LF4)
LF 5
• Use the keys
APPROVAL CONFIRMITY
DISPLAY
nonE Version not approved
LF 6
DON'T MODIFY, IN NO WAY, THIS PARAMETER. IT MUST ALWAYS BE CONFIGURED AT
nonE.
!
The change of this parameter implies the blocking of some functionalities.
GRAVITY
LF 7
*Introduce the gravity of your zone before doing the first calibration.
*Introduce the gravity of destination after doing the calibration.
*The value of the gravity will be denied if it is bigger than 9.83217 ( gravity of the pole) or inferior to 9.78031 ( gravity of the
equator).
Value of factory: 9.8035
DISPLAY
and then the key
to exit from the configuration and go back to the menu LF4)
,
and then
to exit from the menu and the scale will start again automatically.
DESCRIPTION AND SEQUENCE OF USE
DESCRIPTION AND SEQUENCE OF USE
to start or the key
and the key
,
and then the key
to exit from the menu and the scale is going to start automatically.
LF 6
DESCRIPTION AND SEQUENCE OF USE
DESCRIPTION AND SEQUENCE OF USE
to select the speed of the AD converter ( press the key
To continue with the other adjustment and press the key
to exit from the menu and the scale will start again
to select how many divisions will show when in zero ( press
to continue with other adjustment or press the key
USER'S MANUAL BR16
EN
63

Publicidad

Tabla de contenido
loading

Tabla de contenido