Coltene CanalPro CL Instrucciones Para El Uso página 41

Pieza de mano eléctrica inalámbrica
Ocultar thumbs Ver también para CanalPro CL:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 75
„ Auto Reverse Function
Auto reverse setting can be changed by pressing the SELECT key for one second or longer while the motor
handpiece stops. As the auto reverse mark flashes while the SELECT key is kept pressed, press the UP/
DOWN key to adjust it.
AUTO REVERSE
AUTO STOP
Auto Reverse OFF
The alarm will sound when it becomes a value of half of a torque limit set up during motor handpiece rota-
tion, and the alarm will change near the torque limit value. (Load alarm) When it continues applying a load
and it exceeds the torque limit value you can select mode from the following three modes.
AUTO REVERSING
The handpiece stop that it reaches for the loaded value, the file rotates in reverse. When a load is removed,
the file returns to normal rotation (forward rotation) automatically
Load lower than
the set torque limit value
Forward rotation
AUTO STOP
The motor handpiece starts in reverse. When a load is removed, the motor handpiece stops. If you want it
to rotate (forward-rotate) again, re-press the ON/OFF Key or re-step.
Load lower than
the set torque limit value
Forward rotation
(flashes while the SELECT key is kept pressed, and goes out when it is
released)
Load higher than
the set torque limit value
Reverse rotation
Load higher than
the set torque limit value
Reverse rotation
Reverse rotation when load continues,
forward rotation when load is removed
Returns to forward
Reverse rotation
when load continues,
stop when load is removed
Stop
Fig. 9
Fig. 10
41

Publicidad

Tabla de contenido
loading

Tabla de contenido