Guidance and manufacturer's declaration – electromagnetic immunity
The NTF3000AR is intended for use in the electromagnetic environment
specified below. The customer or the user of the NTF3000AR should ensure
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Radiated RF
IEC 61000-4-3
Conducted RF
IEC 61000-4-6
Electrical fast transient
IEC 61000-4-4
Surge
IEC 61000-4-5
Power frequency
magnetic field
IEC 61000-4-8
Voltage dips, short
interruptions and
voltage variations on
power
supply input lines
IEC 61000-4-11
12
that it is used in such an environment.
IEC 60601 test
Compliance level
level
±6kV Contact
±8kV Air
3 V/m 80MHz to
2.5GHz
3Vrms 150kHz to
80MHz
±2kV power line
±1kV I/O lines
±1kV
differential
±2kV common
3 A/m
>95% dip
0.5 cycle
60% dip 5 cycles
70% dip 25 cycles
95% dip 5 sec.
Electromagnetic environment
concrete, or ceramic tile. If floors
Complies
material, the relative humidity
shielded location from fixed RF
Complies
transmitters, as determined by
an electromagnetic site survey,
Interference may occur in the
vicinity of equipment marked
Not Applicable
(no electrical
cabling)
Separation distance calculation
calculated using the equations.
Not Applicable
Not Applicable
Complies
Not Applicable
– guidance
Floors should be wood,
are covered with synthetic
should be at least 30%.
Field strengths outside the
should be less than 3 V/m.
with the following symbol:
provided below. If a known
transmitter is present the
specific distance can be
The ME equipment
is solely battery powered.
Power frequency
magnetic fields should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.
The ME equipment
is solely battery
powered.