Guidance and manufacturer's declaration – electromagnetic emissions
The NTF3000AR is intended for use in the electromagnetic environment
specified below. The customer or the user of the NTF3000AR should ensure
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Non-Life Support Equipment Separation Distance Calculation
Rated maximum output
power of transmitter (W)
0.01
0.1
1
10
100
that it is used in such an environment.
Compliance
The ME equipment uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
Group 1
Class B
Not
Applicable
Not
Applicable
(3Vrms / 3V/m compliance)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
in ISM bands
3,5
d = [
] P
V
1
0.12
0.37
1.17
3.69
11.67
Electromagnetic environment – guidance
are not likely to cause any interference in nearby
electronic equipment.
Complies
The ME equipment is
solely battery powered.
80 MHz to
800 MHz
3,5
d = [
] P
E
1
0.12
0.37
1.17
3.69
11.67
800 MHz to 2.5 GHz
7
d = [
] P
E
1
0.23
0.74
2.33
7.38
23.33
13