Setup Menu: Configuration Menu - ResMed Stellar 100 Guia Del Usuario

Ocultar thumbs Ver también para Stellar 100:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Parameter
Learn Circuit
FiO
Sensor
2
Calibration
Ramp Time

Setup menu: Configuration Menu

Parameter
Language
Brightness
Backlight
Time format
Date format
To return to the Options screen press the Return button
Description
The device is calibrated according to your air tubing system.
1. If in use, turn off the oxygen flow.
2. Select the mask type.
3. Set up the air circuit including accessories and patient interface.
Note: For invasive use, do not connect a catheter mount,
tracheostomy tube or HMEF, see "Setting up for invasive use"
on page 5).
4. Leave the air circuit unobstructed and open to the air.
5. Press
to start the Learn Circuit.
6. Wait for the device to complete its automated tests (<30 sec).
The results are displayed when complete. If the circuit
configuration has been successfully learnt,
unsuccessful,
displays (see "Troubleshooting" on page 21).
The device starts the calibration of the FiO
measure the oxygen concentration of the breathable air.
1. Press
to start the FiO
2. Wait for the device to complete its calibration.
The results are displayed when complete.
Note: Turn off the oxygen flow.
If Max Ramp Time has been set by the clinician, you may select any
value up to this time.
Options: 0 min-Max Ramp Time (max 45 minutes, 5 minutes
increments)
Description
Sets the display language.
Options: Depending on regional configuration
Sets the LCD backlight brightness.
Options: 20-100%, 10% increments
Enables the LCD and keypad backlight.
If the AUTO setting is selected the backlight turns dark after five
minutes without any actions and turns on again if any button is
pressed again or an alarm occurs.
Options: On, Auto
Sets the time format.
Options: 24 hrs,12 hrs
Sets the date format.
Options: dd/mm/yyyy, mm/dd/yyyy
displays. If
monitoring sensor to
2
sensor calibration.
2
Setup menu
17

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Stellar 150

Tabla de contenido