LP-30-220 • LEEP PRECISION
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The CooperSurgical LEEP PRECISION Smoke Evacuation Systems are
intended for use in the electromagnetic environment specified below. The
customer or the end user of the CooperSurgical LEEP PRECISION Smoke
Evacuation System should assure that it is used in such an environment.
Immunity Test
Electromagnetic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level; in this case, 230 V.
T
Operating Manual • English (continued)
IEC 60601
Test Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3 A/m
TM
Smoke Evacuation System, 230 VAC
Compliance
Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common
mode
<5% U
T
(>95% dip in U
for 0.5 cycle
40% U
T
(60% dip in U
for 5 cycles
70% U
T
(30% dip in U
for 25 cycles
<5% U
T
(>95% dip in U
for 5 sec
3 A/m
8
Electromagnetic
Environmental -
Guidance
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 percent.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
)
T
commercial or hospital
environment.
If the user of the
)
T
CooperSurgical LEEP
PRECISION Smoke
Evacuation System
requires continued
)
operation during power
T
mains interruptions, it is
recommended that the
CooperSurgical LEEP
PRECISION Smoke
)
T
Evacuation System be
powered from an
uninterruptible power
supply or a battery.
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
typical commercial or
hospital environment.