The WHO classification in the display shows the range of
the measured blood pressure.
If the values for systolic and diastolic pressure are in two
different WHO ranges (e.g. systolic in the grade hyperten-
sion range and diastolic pressure in the normal range), the
WHO classification on the unit always indicates the higher
range (grade hypertension in the example described).

7. Saving, retrieving and deleting results

together with date and time. With more than 60 items of
measured data, the earliest items of data measured are
lost.
position.
firm the desired user memory with the "START/STOP"
button. If you press the "MEM" button again, the average
value " " of all the stored measured values in the user
memory will be displayed. If you press the "MEM" button
again, the average value of the last 7 days for the morning
measurement will be displayed (Morning: 5 a.m. - 9 a.m.,
display "
"). If you press the "MEM" button again, the
average value of the last 7 days for the evening measure-
ment will be displayed (Evening: 6 p.m. - 8 p.m., display
"
"). If you continue to press the "MEM" button, the
most recent individual measured values with date and
time are displayed in turn.
"MEM" button for 3 seconds. All the values in the current
user memory are deleted after three beeps are output.
or the "START/STOP" button or push the touch screen
activation switch to "OFF".
matically after 2 minutes.

8. Error messages/trouble-shooting

In case of faults, the "Er_" message appears in the display.
Error messages can occur when
the blood-pressure values are unusually high or low (Er1
or Er2 appears in the display),
the cuff is too tight or too loose (Er3 or Er4 appears in the
-
display),
the inflation pressure is higher than 300 mmHg (Er5 ap-
pears in the display),
inflation takes longer than 3 minutes (Er6 appears in the
display),
there is a system or device error (ErA, Er0, Er7 or Er8 ap-
pears in the display),
the batteries are almost empty
In such cases, repeat the measurement. Ensure you do not
move or speak.
If necessary, re-insert or replace the batteries.