2) Confirm Yes to replace or No to cancel. If Yes is selected the system will begin scanning
for the new disposable.
3) Prime the disposable DO NOT PULL BATTERY TAB OF DISPOSABLE UNTIL
PRIMING IS COMPLETED. System will continue scanning until a connection is made.
4) Pull battery tab of new disposable and confirm wireless connection.
BLUETOOTH DISCONNECT EVENTS
If the disposable device temporarily loses Bluetooth connection, the system will automatically
attempt to reconnect to the disposable. Contrast monitoring will be suspended while the system
attempts to reconnect and the user is notified of the disconnect.
TEMPORARY LOST CONNECTION SCREEN
If the disposable device permanently loses Bluetooth connection and the connection can not be
re-established, contrast accounting will continue to be suspended and the user is notified of the
disconnect event.
The user has the option to end the case or replace the disposable. If the disposable is not
replaced and DyeVert Plus or DyeVert Plus EZ is being used the case may continue with contrast
savings but without contrast accounting. If the disposable is not replaced when a connection
cannot be re-established, the user can manually track remaining injections and add the
additional contrast amount injected to the cumulative total.
PERMANENT DISCONNECT SCREEN
Possible Alarm Messages
Alarm Message
Threshold
%
Alarm
and
Threshold
Reached Alarm
Low Battery Alarm (Disposables)
No Battery Alarm (Disposables)
Possible Error Messages
Error Message
"The connected Syringe is not calibrated
properly. Please connect another device
to continue."
"The connected DyeVert is not calibrated
properly. Please connect another device
to continue."
"No syringe or DyeVert data received."
"No syringe or pressure module data
received."
"The
syringe
cannot
be
connected.
Replace the disposable and continue the
case, or end the case?"
"The
syringe
cannot
be
connected.
Replace the disposable and continue the
case, or return to main menu?"
"The DyeVert cannot be connected.
Replace the disposable and continue the
case, or end the case?"
"The DyeVert cannot be connected.
Replace the disposable and continue the
case, or return to main menu?"
"The
Pressure
Module
cannot
be
connected. Replace the disposable and
continue the case, or end the case?"
"The
Pressure
Module
cannot
be
connected. Replace the disposable and
continue the case, or return to main
menu?"
"There is no remaining battery power for
the syringe."
"There is no remaining battery power for
the DyeVert."
"There is no remaining battery power for
the Pressure Module."
"Would you like to continue the current
case started on <date> at <time>?"
"Unable to connect to the <device name>.
Please replace the unit."
"Please wait as we reconnect to previously
used devices."
"Please wait as we dispose of previously
used devices."
"Database
Error:
failed
to
initialize
database."
"Database Error: failed to add case."
"Database Error: failed to update case."
"Database Error: failed to delete all
cases."
"Case Upload Failed"
"Passcode incorrect. Please try again."
"Smart Monitor Not Charging: This Smart
Monitor is not connected to a power
source"
User Actions
Notify physician that pre-set threshold volume has
been reached. This alarm is for notification
purposes only if warning tones or voice alerts are
enabled.
If warning tones or voice alerts are enabled, alarm
will sound for approximately 2 seconds. User must
replace the disposable, end the case or allow the
battery to continue depleting while continuing the
case. Monitoring capability will continue until the No
Battery Alarm sounds.
User must replace the disposable to continue
contrast accounting or continue case without
monitoring capability. If warning tones or voice
alerts are enabled an audible alert will sound in
addition to the warning on the screen.
User Actions
Replace the Smart Syringe.
Replace the DyeVert Plus or DyeVert Plus EZ
Module disposable.
End case and replace the Smart Syringe and
DyeVert Plus or DyeVert Plus EZ Module
disposable.
End case and replace the Smart Syringe and
Pressure Module disposable.
Replace the Smart Syringe or end case.
Replace the Smart Syringe and continue the case or
select main menu and cancel device use.
Replace the DyeVert Plus or DyeVert Plus EZ
Module disposable or end case.
Replace the Dyevert Plus or DyeVert Plus EZ
Module and continue the case or select main menu
and cancel device use.
Replace the Pressure Module disposable or end
case.
Replace the DyeTect Pressure Module and continue
the case or select main menu and cancel device
use.
Replace the Smart Syringe.
Replace the DyeVert Plus or DyeVert Plus EZ
Module disposable.
Replace the DyeTect Pressure Module disposable.
Select "Yes" to restart the case, "No" to end the
case.
Replace Osprey Medical disposable device.
No action required.
No action required.
Reboot Smart Monitor by pressing and holding
Home button and top Sleep/Wake button until
screen turns off and keep holding down buttons until
screen turns on again. If the problem persists,
contact Osprey Medical.
Contact Osprey Medical.
Contact Osprey Medical.
Contact Osprey Medical.
Contact Osprey Medical.
Retry passcode entry. If the problem persists,
contact Osprey Medical.
Plug in Smart Monitor to the supplied Lightning
Cable and power supply.
8280-B