Cleaning the Transducer

The transducer requires no routine maintenance or servicing.
To sterilise or clean the transducer it may be removed by means of
the following procedure:
1
Remove the transducer by gently rotating the transducer
anti-clockwise and then pulling from the main body.
2
The transducer may now be immersed in warm soapy water
for routine cleaning or immersed in cold sterilising solutions
e.g. Perasafe for a period not exceeding 10 minutes. (Alcohol
and chloride solutions should be avoided.) After cleaning/
sterilising, the transducer should be rinsed in distilled water
and dried. Perasafe is available from Vyaire in convenient
81g plastic bottles, Cat No.36-SSC5000A.
3
Re-assemble the transducer into the Micro I.
CAUTION: Do not attempt to wash or immerse the spirometer
or accessories in water or cleaning fluid, as there are electronic
components inside that will be permanently damaged.

Servicing

If your unit requires service or repair please see page 288 for contact
details.
There are no user serviceable parts in the Micro I.

Product Lifetime

The Micro I meter is designed for a product lifetime of 5 years.