Electromagnetic Compatibility (Emc); To Iec 60601-1-2 - Vyaire Micro I Manual De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 172

Electromagnetic Compatibility (EMC)

to IEC 60601-1-2

WARNING: Use of portable phones or other radio frequency
(RF) emitting equipment near the system (<30cm) may cause
unexpected or adverse operation.
The Micro I has been tested to IEC 60601-1-2:2014, regarding its ability to
operate in an environment containing other electrical/electronic equipment
(including other medical devices).
The purpose of this testing is to ensure that the Micro I is not likely to
adversely affect the normal operation of other such equipment and that other
such equipment is not likely to adversely affect the normal operation of the
Micro I.
Despite the testing of the Micro I that has been undertaken, normal operation
of the Micro I can be affected by other electrical/electronic equipment and
portable and mobile RF communications equipment. Keep a distance of about
2 meters from possible error sources when using the device.
As the Micro I is medical equipment, special precautions are needed regarding
EMC (electromagnetic compatibility).
It is important that the Micro I is configured and installed/put into service, in
accordance with the instructions/guidance provided herein and is used only in
the configuration as supplied.
Changes or modifications to the Micro I may result in increased emissions or
decreased immunity of the Micro I in relation to EMC performance.
The Micro I should be used only with the accessories (USB cables, mains
adapter and turbine transducer) supplied (which are referenced in the
accessories section of this manual). None of the Micro I cables should be
extended in length by the user.
If any cables are extended by the user or non approved accessories are
used, this may result in an increased level of emissions or decreased level
of immunity, in relation to the Micro I EMC. None of the Micro I accessories
should be used with other devices, as this may result in an increased level
of emissions or decreased level of immunity, in relation to the other devices'
EMC.
List of EMC important cables. Do not exchange by other types.
Part Number
36-CAB1098
Description
USB Cable (Micro I to PC or to PSU)
40