Guidance and Manufacturer's Declaration – Electromagnetic
Immunity
The Micro I is intended for use in the electromagnetic environment specified
below. The customer or the user of the Micro I should assure that it is used
in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
IEC 60601 test
level
Contact: ± 8 kV
Air:
±2 kV, ± 4 kV, ±
8 kV, ± 15 kV
± 2 kV
100 kHz
repetition
frequency for
power supply
lines
± 0,5 kV, ± 1 kV,
± 2 kV
0 % U
; 0,5
T
cycle
At 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0 % U
; 1 cycle
T
and
70 % U
; 25/30
T
cycles
Single phase:
at 0°
and
0 % U
; 250/300
T
cycle
Compliance
Level
Contact: ± 8 kV
Air:
±2 kV, ± 4 kV, ±
8 kV, ± 15 kV
± 2 kV
100 kHz
repetition
frequency for
power supply
lines
± 0,5 kV, ± 1 kV,
± 2 kV
0 % U
; 0,5 cycle
T
At 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0 % U
; 1 cycle
T
and
70 % U
; 25/30
T
cycles
Single phase:
at 0°
and
0 % U
; 250/300
T
cycle
42
Electromagnetic
Environment
-Guidance
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Mains power
quality should be
that of a typical
commercial
or hospital
environment. If the
user of the Micro I
requires continued
operation during
power main
interruptions, it
is recommended
that the Micro I
be powered from
an uninterruptible
power supply or a
battery.