NOMENCLATURE
PENDANT FAMILIARIZATION
Fast Back™
Carriage will slow down within 15° of
the flexion angle.
Mode
Used with Fast Back, Oscillation Zone,
and Progressive ROM*
Reset
Clear previous treatment
settings and completed
cycles.
Extension
Adjusts the extension angle.
Flexion
Adjusts the flexion angle.
Speed
Adjusts the speed of operation.
Extension/Flexion Delay
Allows user to program carriage to stop at both the
extension and flexion angle for the time programmed.
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
Oscillation Zone™
Carriage will oscillate three times between
the flexion angle and 10° less than
the flexion angle.
6
Progressive ROM™ *
Unit will automatically increase the
programmed flexion angle by 1° every hour
up to a maximum of 5° per day.
Up Arrow
Allows user to increase
treatment parameters.
Down Arrow
Allows user to decrease
treatment parameters.
Comfort Zone™
Temporarily decreases the flexion ROM*
when discomfort is experienced.
Start/Stop
(Emergency Stop)
Start or Stop Treatment.
* Range of Motion