5 HYGIENIC RE-USE IN THE PROFESSIONAL
ENVIRONMENT

5.1 General

Ensure that the system is processed ...
NOTICE
Ensure that the system is processed for re-use only with the
application of adequately validated device- and product-specific
procedures, and that the validated parameters are maintained
during each cycle. The equipment used must be maintained
regularly.
Follow the schedule defined below for hygienic re-use:
– Dispose of the used filter pad after every inhalation session.
– Clean and disinfect the PARI Filter/Valve Set after every inha-
lation session.
– Clean, disinfect and sterilise the PARI Filter/Valve Set before
every change in patient.
The PARI Filter/Valve Set must be processed for re-use with the
application of suitable, validated methods in such manner that
the success of these methods can be monitored reliably and the
safety and health of patients is not threatened.
Ensure that the components are allowed ...
CAUTION
Ensure that the components are allowed to dry sufficiently after
each processing step. Condensation or residual wetness can
present an increased risk through the growth of bacteria.
The following text describes the methods validated by PARI, with
which plastic PARI parts can be processed effectively.
Ensure that the desired results are achieved consistently by the
Ensure that the desired results are achieved ...
processing methods employed by your staff with the equipment
and chemicals used. Take particular care to ensure that the
- 34 -