processing procedure you select is effective if it must differ from
our validated procedures (e.g., because the equipment and
chemicals described are not available). When selecting the
chemicals for use, make sure that they are suitable for cleaning
or disinfecting medical products made from the materials indi-
cated [See also: 2.2 Material information]. Bear in mind the man-
ufacturer's dosing information and instructions for safe handling
of chemicals.
To assist you in processing the system for hygienic re-use, a
checklist may be obtained from PARI GmbH (on the internet at
www.pari.de/en/products or from the PARI Service Center [See
also: 6.2 Contact]).

5.2 Processing limits

The PARI Filter/Valve Set can be sterilised up to 300 times and
used for up to one year.
Do not expose the PARI Filter/Valve Set to temperatures above
137 °C.

5.3 Preparation

Dismantle the PARI Filter/Valve Set into its individual parts:
• Open the exhalation filter (bayonet closure).
• Take the filter pad out and dispose of it.

5.4 Cleaning and disinfection

Clean and disinfect immediately after use. Ideally, a mechanical
procedure (instrument dishwasher) is to be used.
When selecting the cleaning agent ...
NOTICE
When selecting the cleaning agent and/or disinfectant, ensure
that it is compatible with the materials used, otherwise damage
to the product components cannot be ruled out.