Prodigy Pocket El Manual Del Propietario página 18

Ocultar thumbs Ver también para Pocket:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

PRODIGY Pocket® Blood Glucose Meter
Questionable Results:
If the test result falls outside the range printed
on the test strip vial, check the "Trouble-Shooting
Guide" located in the "System Troubleshooting"
section of this manual and repeat the test.
Results that fall outside of the expected control
ranges may indicate:
• An error in performing the test
• Control solution bottle not shaken well
• Expired or contaminated control solution
• Control solution that is too warm or too cold
• Test strip deterioration
• Meter malfunction
• Test strips expired
Owner's Manual
PRODIGY Pocket® Blood Glucose Meter
Caution: Do not use test strips or control
solution that have exceeded the discard date,
are expired or have been damaged.
Your results may be inaccurate.
Contact Customer Care
If your test strips have been damaged or if you
continue to get out-of-range results, it means
that the system or the control solution may not
be working properly. DO NOT use the system to
test your blood glucose level. If you are unable to
resolve the problem, contact Customer Care at
1.800.243.2636.
34
Owner's Manual
35

Publicidad

Capítulos

Tabla de contenido
loading

Productos relacionados para Prodigy Pocket

Tabla de contenido