SPECIFICATIONS
MECHANICAL
Height:
Width:
Depth:
Weight:
OPERATIONAL
Flow Rate:
Dose:
Increments:
Priming Rate:
Accuracy:
Occlusion Detection
Pressure:
Head height effect:
Operating
Temperature:
Operating Humidity:
Atmospheric pressure:
Drop
Programming Modes:
Operating Mode:
Alarms:
Infusion sets, models
and installation:
Page 60
140 mm
120 mm
80 mm
490 g
1 mL/hr to 400 mL/hr
1 mL to 9999 mL (feed until empty when DOSIS (SET DOSE) at INF DOSIS
(infinite dose)
1 mL
≥ 700 mL/hr
± 5%, under recommended head height (see PUMP ACCURACY TESTING)
Upstream:
Downstream:
≤ 2.5% per 30 cm (12 inches) (see PUMP ACCURACY TESTING)
+5° to +40° C
10% to 95% non-condensing
65 kPa to 102 kPa
Dropping from any angle from a height of 1 meter (3 feet) should not
damage pump operation.
SET RATE, SET DOSE (See INSTRUCTIONS FOR USE)
RUN
See ALARM AND WARNING MESSAGES
See FEEDING SETS
44838 Rev. J
-34 ± 20 kPa
103 ± 20 kPa for 30 seconds
124 ± 20 kPa for 5 seconds