Gima H100 Instrucciones Para El Uso página 74

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Switch the green cut-out on (see Picture no. 17)
Turn the knob clockwise completely (15 minutes)
Now the automatic sterilization cycle
"heating" light (4) turn on. About two minutes after the pressure gauge (1) rises up (picture no. 9 on page
no. 70) and from this moment on the lid cannot be opened any more.
When the pressure reaches the one corresponding to 100 °C the working valve rises up and closes the air
bleeder.
At this moment the autoclave stats to rises its steam. When the temperature reaches 125 °C the real sterilizing
cycle begins and therefore the green "sterilizing" gauge lights on (5). Now the programmer knob (2) begins
automatically to turn conter-clockwise (see picture no. 19).
.
Autoclave GIMA H100 automatic
0
START
STERILIZING
COMPLETE
Picture no. 17
0
E
15'
Picture no. 18
begins and the green START light (3) togheter with the
HEATING
STERILIZING
ALARM
POWER
START
HEATING
STERILIZING
STERILIZING
ALARM
COMPLETE
POWER
- 70 -

Publicidad

Capítulos

Tabla de contenido
loading

Productos relacionados para Gima H100

Tabla de contenido