DISTILLED WATER TOPPING-UP
Before starting a new cycle it is necessary to check the distilled water level. Make sure that the tempera-
ture control probe is always and completely covered with water
Warning :
!
When filling up the water please follow carefully the above mentioned instructions. In any
case do not fill up over the upper hole (FS) of the chamber (see picture no. 14 on page
67).
USED WATER DISCHARGE
Every 10 days, or after 5/10 cycles, empty the autoclave completely and incline it on one side
no. 23).
Autoclave GIMA H100 automatic
Picture no. 23
- 73 -
.
(see picture