Warning:
!
Do not wash the autoclave with direct or high pressure water jets.Any
electrical or electronic components could damage the regular working of the unit and its safety
system.
CLEANING OF THE STERILIZING CHAMBER, THE BASKET AND THE
HOLDER
Accurately clean (every week) the sterilizing chamber, the basket and the holder with a
non-abrasive cloth and water or cleansing for stainless steel objects. Then rinse with alcohol and
later with water.
Warning:
!
Never use disinfectants when cleaning the chamber.
Important :
Only use distilled water, otherwise calcareous deposits may obstruct valves..
SUBSTITUTION OF THE GASKET
Open the lid and remove the gasket from its seat. Clean the seat of the gasket with a cloth and alcohol. Insert
the new one, after pouring some talcum on it, in the seat of the lid.
CLEANING OF THE WORKING AND SIGNAL VALVES
Regularly check that the working valve is always well cleaned and rises when achieving a 100°C tempera-
ture
!
Warning:
The rare emission of water drops and a light venthole during the sterilizing cycle means that the valve is
working correctly.
-
Check if the pressure signal is clean and runs freely in its seat.
REPLACEMENT OF THE FUSES
Unplug the power cable from the main power, move the power cable from the small container situated on the back
side of the autoclave (see detail no. 8, picture no. 11 on page 65). With a small screwdriver unscrew the small
drawer containing the fuses (detail no. 9, picture no. 11 on page no. 65). Remove the fuses and replace both of
them with two of the same kind and value. To determine their value see on the serial label (detail no. 7, picture no.
11 on page no. 65).
Autoclave GIMA H100 automatic
- 79 -
infiltrations inside