Preventive Maintenance - Gima H100 Instrucciones Para El Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

PREVENTIVE MAINTENANCE

Warning !
!
Before making any maintenance, make sure that the line
connected, in order to avoi electrical
EVERY 10 DAYS OF
AFTER 5/6 CYCLES
MONTHLY
EVERY SIX MONTHS
EVERY 2 YEARS
CLEANING OF THE LID GASKET
Clean either the gasket or the stainless steel lid with a cloth and water or vinegar to
nate calcareous scales. Use a non-abrasive wet cloth and normal cleansing for stainless steel to
clean mirror side od the lid.
Warning:
!
Do not accumulate calcareous or dirt residues on the gasket because they can cause its
damaging or even breaking.
CLEANING AND DISINFECTION OF EXTERNAL SURFACES
Periodically clean every external sides of the unit using a wet cloth and normal cleansing or
simply water.
Use either denatured alcohol or cleansing with few hypoclorited sodium (or similar) to disinfect
external sides.
Autoclave GIMA H100 automatic
DAILY
WEEKLY
YEARLY
cord of the autoclave is not
shocks or contacts with steam.
Clean the gasket and the lid
Clean the external surfaces
Clean the internal surfaces
Clean the sterilizing chamber
Clean the basket and its bearing
Clean and disinfect external surfaces
Empty the autoclave completely inclining it on one side,
then pour some distilled water and cover the
temperature
control probe with water.
maintenance
of the working and pressure valves
check sterilizing effectiveness with the biological
indicators
complete control by the technical assistance
substitution of the lid gasket
- 78 -
elimi

Publicidad

Capítulos

Tabla de contenido
loading

Productos relacionados para Gima H100

Tabla de contenido