ELECTROMAGNETIC COMPATIBILITY
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
This monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of this monitor should assure that it is used in such an environment.
Immunity Test
Conducted RF
3 Vrms 150 kHz to
IEC 61000-4-6
80 MHz
Radiated RF
3 V/m 80 MHz to
IEC 61000-4-3
2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which this monitor is used exceeds the applicable RF compliance level above, this monitor should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating this monitor.
** Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
30 • ENGLISH
IEC 60601 Test Level
3 Vrms
3 V/m
Table 3
Compliance Level
Electromagnetic Environment Guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of this monitor, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d =
d =
d = 2.3
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,*
should be less than the compliance level in each
frequency range.**
Interference may occur in the vicinity of
equipment marked with the following
symbol:
1
2 .
80 MHz to 800 MHz
2 .
1
800 MHz to 2,5 GHz