6 Preparation for use
WARNINGS
For use only by medically trained persons who have been adequately trained in suction
procedures and in the use of aspirators.
CAUTIONS
The Basic 30 must remain in an upright position during use.
Sterile accessories must be checked on the integrity of the packaging before use.
Non sterile and reusable accessories must be cleaned, disinfected and / or sterilized
according to the Medela cleaning guide (product code 200.2391).
6.1

Checks before use

6.2
Assembly of the
basic configuration
14 |
Preparation for use
Check the Basic 30 system before use for damage of
the power cord or plug, obvious device damage or
safety defects and proper functioning of the device.
Check for completeness and general condition of the
Basic 30 delivery package.
Check all accessories prior to use:
– Suction jars, lids and liners for cracks, brittle and
flawed spots. Replace if necessary.
– Tubing for cracks, brittle areas and that connectors
are firmly attached. Replace if necessary.
– As an additional safety test, evacuate the system
(including jars) to maximum vacuum before actual
use.
1.
2.
3. Attach all necessary accessories according to your needs.
See chapters 17 / 18 for more details.
1.1 Make sure that the Safety
Set is attached to the
Basic 30. See chapter 5.5.
2.1 Attach a filter (if required)
to the safety set with the
arrow pointing in the flow
direction. The filter must
be exchanged after an
overflow and after each
patient, but latest after a
week.