11 Troubleshooting

Motor not running

Check if:
the Basic 30 is switched on. The standby LED must be
illuminated.
the mains voltage is correct and if the mains plug is
inserted correctly into the fixed mains socket.
the fuse on the back of the Basic 30 is not defective. For replacing the defective fuse see
chapter 12.
If the fault cannot be rectified, please contact the internal technical department.

Insufficient vacuum

Check if:
the vacuum regulator is set correctly.
the tubings are not defective or broken. If necessary, replace.
all plug-in connections are tight.
the overflow protection is deactivated / open. If the overflow protection is activated,
deactivate as shown under 6.4a/2.2.
the suction jar / lid have no cracks, brittle areas, discolouration. If necessary, replace.
the disposable system has no cracks, brittle areas,
discolouration. If necessary, replace.
filter is not clogged. To test if the filter is clogged see
chapter 13.
If the fault cannot be rectified, please contact the internal technical department.

No lED lit

The Basic 30 is not connected to the mains or the fuse needs replacement.
22 |
Troubleshooting