22 Technical documentation (EMC)
WARNINGS
Do not use other accessories than those specified or sold by the manufacturer as
replacement parts for internal components as it may result in increased emissions or
decreased immunity of the Basic 30 pump.
EMC
Basic 30 is EMC-tested in conformity with the requirements of IEC 60601-1-2:2007 and can
be used in the vicinity of other EMC-tested devices that fulfill the requirements of the relevant
IEC 60601-1-2 standard. Untested HF (high-frequency) sources, radio networks or the like
can influence the operation of the device and may not be operated in combination with the
system. Basic 30 is a medical device that requires special safety precautions and must be
installed and placed in operation in accordance with the attached EMC information. Portable
and mobile RF communication devices (mobile telephones) can affect Basic 30.
Electromagnetic Emissions
Basic 30 is intended for use in the electromagnetic environment specified below. The
customer or the user of Basic 30 should assure that they are used in such environ-
ment.
Emission Tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctutions /
flicker emissions
IEC 60000-3-3
WARNINGS
Basic 30 should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, Basic 30 should be observed to verify normal operation in the
configuration in which it will be used.
382 |
Technical documentation (EMC)
Compliance
Electromagnetic environment – guidance
Group 1
Basic 30 uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
with nearby electronic equipment.
Class B
Basic 30 is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage
Class A
power supply network that supplies buildings
used for domestic purposes.
Complies