Debriefing After Use - Laerdal CPRmeter Instrucciones De Uso

Ocultar thumbs Ver también para CPRmeter:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 79

5 Debriefing After Use

Q-CPR
®
The CPRmeter can display CPR performance statistics for the last CPR
event. When the CPRmeter is turned on, press the On / Off button once to
activate Q-CPR Quick Review. The statistics are shown over two displays.
Percentage of compressions with release force
within the Compression Release Target.
Percentage of compressions with compression
depth which have reached the Compression
Depth Target.
Percentage of compressions with compression
rate within the Compression Rate Target.
Press the On / Off button once to cycle to the next
display.
Duration of the CPR event (minutes: seconds).
Percentage of the CPR event duration where
chest compressions have been delivered.
Press the On / Off button once to cycle between
the displays.
To turn the CPRmeter off, press and hold the On / Off button for at least 1
second.
The CPRmeter reverts to Compression Feedback mode if a compression is
delivered.
The CPR event statistics are stored when the CPRmeter is turned off. When
turned on again, the statistics from the stored CPR event can be reviewed,
as described above.
When the CPRmeter is used in a new CPR event, the preceding event's
statistics are deleted and the new event's statistics are stored in Q-CPR
Quick Review.
CPR performance statistics are only calculated if at least 10 compressions
have been delivered.
CPRmeter_DFU_N2131_RevE.indd 15
Quick Review
15
98%
70%
96%
24:35
73%
19.03.12 14:54

Publicidad

Tabla de contenido
loading

Tabla de contenido