RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER -
Rapid electrical transients in
Voltage trough, short power cuts
and voltage variations on power
NOTE: VT is the AC supply voltage before application of the test
level.
178
Compex Energy is designed for use in the electromagnetic environment
The buyer or user of the Compex Energy must ensure it is used
Immunity test
Electrostatic discharge
(DES)
CEI 61000-4-2
bursts of noise
CEI 61000-4-4
Surge currents
CEI 61000-4-5
supply input lines
CEI 61000-4-11
Magnetic field
at mains frequency
(50/60 Hz)
CEI 61000-4-8
stipulated below.
in this recommended environment.
Test level IEC 60601
± 6 kV at the contact
± 8 kV in air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5% VT (trough > 95% of UT)
for 0.5 cycle
< 40% VT (trough > 60% of VT)
for 5 cycles
< 70% VT (trough > 30% of VT)
for 25 cycles
< 5% VT (trough > 95% of VT)
for 5 seconds
3 A/m
Observance level
± 6 kV at the contact
± 8 kV in air
Not applicable
System battery-powered
Not applicable
System battery-powered
Not applicable
System battery-powered