RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER -
Compex Energy is designed for use in the electromagnetic environment stipulated
a
The field intensity from fixed transmitters, such as radio telephone base stations (cellular/wireless) and a
mobile radio, amateur radios, AM and FM radio emissions and TV emissions cannot be predicted with any
accuracy. It may therefore be necessary to consider an analysis of the electromagnetic environment of the
site to calculate the electromagnetic environment coming from RF fixed transmitters. If the field intensity
measured in the environment where the Compex Energy is located exceeds the appropriate RF observance
NOTE: VT is the AC supply voltage before application of the test
level.
180
below. The buyer or user of the Compex Energy must ensure it is used in this
Immunity test
RF immunity
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: From 80 MHz to 800 MHz, the spacing for high frequency amplitude is applied.
NOTE 2: These guidelines may not be appropriate for some situations.
level above, the Compex Energy should be monitored to ensure it is operating properly.
recommended environment.
Test level IEC 60601
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
10 V/m
26 MHz to 1 GHz
Observance level
Signal lines
Not applicable for power supplies
of battery-powered systems
3 A/m
10 V/m