Connecting The Probe To The Dut - R&S RT-ZC10 Manual Del Usuario

Tabla de contenido

Publicidad

®
R&S
RT-ZC10
4. Close the clamp until it snaps in.
5. Press the slider on the sensor head until the UNLOCK indication disappears,
and hold it until LOCK appears. Check that the opening lever is firmly locked
and the clamp is securely closed.
6. Press the demagnetizing switch (DEMAG) on the terminator.
7. Turn the zero adjustment dial on the terminator to adjust the trace to the zero
position.
Risk of instrument damage
Make sure to open the clamp of the probe always with the slider.
3.4

Connecting the Probe to the DUT

1. Check that the system is safe, and that the preparations described in the pre-
ceding sections have been carried out.
2. Pull the sensor slider, so that the clamp opens.
3. Align the sensor so that the current direction indication corresponds to the direc-
tion of current flow through the conductor to be measured. The conductor should
be in the center of the clamp aperture because the measurement may be affec-
ted by the position within the clamp aperture of the conductor being measured.
4. Close the clamp until it snaps in.
User Manual 1409.7808.02 ─ 01
Putting into Operation
Connecting the Probe to the DUT
21