II. STARTING UP / INSTALLATION
II. 1.
Use in direct oxygen therapy.
a - Ensure that the switch (1) is in the 0 position.
b  -­  If  used  with  a  humidifier:
Unscrew  the  flask  and  fill  it  with  water  up  to  the  line  (see  the  humidifier  
instructions).  Then  screw  the  humidifier  flask  onto  its  lid  until  there  are  
no leaks from it.
c  -­  Connect  the  oxygen  administration  tube  to  the  humidifier  outlet  
nozzle or connect the administration nasal cannulas onto the
concentrator. The tube between the patient and the NewLife
Intensity should be less than 60 metres long, in order
to  ensure  that  the  oxygen  flow  rate  remains  satisfactory.
d - Ensure that all of the parts are connected correctly so as to avoid
leaks.
e - Plug the power cable into a mains socket.
f - Press the switch to the start position - I . An alarm operating test
is carried out automatically when the machine is switched on (this
test lasts about 5 seconds).
g  -­   T urn  the  flow  adjustment  knob  (6)  to  the  prescribed  value.  
The prescribed valve line should appear to split the middle of the
flowmeter  ball.
h  -­  Check  that  the  oxygen  flows  out  of  the  administration  device  
(nasal  cannulas  or  other)  by  placing  the  orifice(s)  on  the  surface  of  a  
glass  of  water.   T he  flow  should  disturb  the  surface  of  the  water.
i - Adjust the nasal cannulas or mask to suit your face.
Remark: the optimal oxygen concentration is obtained about ten minutes after the device is
switched on (90 % of the concentration is obtained after around 5 minutes).
At the end of the treatment, press the push button (1) to place it in the 0 position to stop the device.
The  oxygen  enriched  air  flow  continues  for  approximately  1  minute  after  the  device  is  stopped.