Do not place DenOptix QST near:
• X-ray equipment that is constantly energized (e.g., devices such as fluoroscopes which are
• Magnetic resonance imaging systems
• Large motor/electric generators
There is potential for electromagnetic interference between devices when using electronic
equipment. International standards exist to minimize this potential.
DenOptix QST has been tested against and complies with the international standard EN
60601-1-2, Medical Electrical Equipment - General Requirements for Safety - Collateral
Standard; Electromagnetic Compatibility Requirements and Tests, class A.
Use only grounded electrical connections.
Connect the DenOptix QST scanner to a grounded electrical outlet between 100-240 volts AC.
Do not reuse the barrier envelopes.
Once used within the oral cavity, the barrier envelopes are bio-contaminated and should
NEVER be reused. Please dispose of in accordance with Section 7.4.
Imaging plates are toxic!
Never place an imaging plate in a patient's mouth without enclosing it first in a completely
sealed barrier envelope. If a patient swallows an imaging plate, contact a physician immedi-
ately. The physician must remove the imaging plate. Do not use cracked, bent or chipped
imaging plates.
Never open or override the lid lock when the scanner is in use or plugged in.
The scanner lid will lock during operation. In case of power failure, unplug the scanner and
open the lid by depressing the lock mechanism as indicated in Section 7.3.
Never use imaging plates with patients that might chew or swallow them.
If the patient bites or chews the plate and damages the protective barrier envelope, rinse the
patient's mouth with a large amount of water. If the patient manages to swallow any of the
blue surface of the plate, contact a doctor immediately.
The DenOptix QST scanner is a Class 1 laser device.
Caution: Use of controls or adjustments or performance of procedures other than those spec-
ified herein may result in hazardous radiation exposure. A fail-safe switch in the carousel well
and cover keeps the laser inactive as long as the cover is open. Only a trained technician from
a qualified Gendex dealer should remove the cover from the scanner. Direct eye contact with
the output beam from the laser may cause serious damage and possible blindness.
User and Service Manual
energized when not in use)