Reading shown is stable
Reading shown is given in grams
Reading shown is given in ounces (1g = 0.03527396200 oz)
Reading shown is given in carats (1g = 5.0000000000 ct)
Reading shown is given in pennyweight (1g = 0.64301493100 dwt)
Reading shown is given in percent weight
Reading shown is given in as a counting
The balance is developing a stable reading
Error operation
The weight on the pan excedes the capacity of the balance
The pan is not properly seated or has been removed
Calibration / Adjustment key
Print / Output key
Count / Function confirmation key
Percentage key
Unit conversion key
Tare key
CALIBRATION
To decide the accuracy of the balance through test the difference between
thereading and the actual weight of the object on the weighing pan.
Prerequisite of calibration:
1. There is no loading on the weighing pan.
2. Press the
3. The interior signals are stable.
Procedure
1. Adjust the level guage and warm it up for
25 minutes.
2. Press the Tare key
3. Press the Calibration key, the display will
read its full range, such as 3000
key.
Key
Display
0.00g
3000