Handpiece; Charger And Handpiece; Transport And Storage Conditions; Installation Of The Unit - 3M Unitek Ortholux Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 49
Classification:
Protection class II,

Handpiece

Power supply:
lithium-ion battery, nominal
voltage 3.7 V
Utilizable wavelength
range:
430-480 nm
Wavelength peak:
455 nm +/- 10 nm
Light intensity (between
400 and 515 nm):
1600 mW/cm
of battery power level)
Light emission area:
typically 42 mm
(optically active)
Intermittent
operation:
1 minute on, 15 minutes off
(ambient temperature of 40°C)
typically 7 minutes operating time
at ambient temperature
(23°C)
Total exposure time
with new, fully
charged battery:
typically 60 minutes
Dimensions:
Diameter: 28 mm
Length: 270 mm
Weight:
250 g (including light guide)

Charger and Handpiece

Time to charge
empty battery:
approximately 2.0 hours
Operating temperature: 10°C to 40°C (59°F to 104°F)
Relative humidity:
30% to 75%
Atmospheric pressure: 700 hPa to 1060 hPa
Total height with
handpiece inserted in
the charger:
180 mm

Transport and Storage Conditions:

Ambient temperature
range:
-20°C to +40°C /
-4°F to +104°F
Relative humidity:
30% to 75%
Atmospheric
pressure:
700 hPa to 1060 hPa
Subject to technical modification without
prior notice.
5
English
All manuals and user guides at all-guides.com

Installation of the Unit

Factory Settings

The factory settings of the unit are as follows:

Initial Steps

Charger

(independent
2
2

Light Guide/Handpiece

Removing and Inserting the Light Guide
from/into the Handpiece
When first turned on the unit is in Extended Mode,
i.e. only the Power level LED is lit on the hand
piece.
Please ensure first that the voltage stated on
the rating plate corresponds to the existing main
supply voltage. The rating plate is attached to the
bottom of the charger.
Place the charger on a level surface.
Connect the power cable of the charger to the
power supply.
- The green LED on the charger showing that the
unit is ready for operation comes on. This shows
that the unit is ready for operation; please refer
to the section "Operating Status Display of the
Charger."
Never place the handpiece in the charger without
the battery inserted in the handpiece!
Place the eye shield on the front of the unit.
Autoclave the light guide prior to first use.
Insert the light guide firmly into the handpiece.
The light guide has a magnetic holder. Remove
the light guide from the handpiece by pulling it
towards the front.
Attach the light guide to the handpiece until it is
firmly in place.

Publicidad