Paragraph 13 – Setting the Pause
Programming
+
PROG
Config.
Paus
N.Open
pH Calibration Menu
You can choose the automatic or manual calibration mode, in both cases the calibration at pH 7 is automatic.
-

Automatic calibration:

When the display shows the value of the buffer solution, insert the probe into the bottle, press
will show the countdown of the 60 seconds required to complete the calibration. If the quality of the alignment is
less than 50%, an error appears on the display; press
automatically exit the calibration); if the quality is higher than 50%, the value appears on the display; pressing
you will be prompted for the pH 4 or 9 buffer solution; at this point the procedure is similar to the previous
calibration.
Manual calibration:
-
When the display shows the value of the buffer solution, insert the probe into the bottle, press
will show the countdown of the 60 seconds required to complete the calibration. If the quality of the alignment is
less than 50%, an error appears on the display; press
automatically exit the calibration); if the quality is higher than 50%, the value appears on the display; pressing
on the display will flash the pH 7.00 value; press
and start the calibration procedure as described previously.
EM00136321
Paus
N.Closed
+
PROG
Config.
Calibr.
Calibr.
Auto.
Probe
Calibr.
Manual
rev.2.0
Remote input to pause the pump.
By default, the system is set to Normally Open.
Press
to access the item, and then press
the value (N. OPEN or N. CLOSED).
Press
to confirm and return to the main menu.
WAIT
Auto Cal
7pH
60 s.
7 pH
Error
7 pH
100% ok
Auto Cal
4/9.25pH
WAIT
60 s.
4 pH
Error
100% ok
4 pH
WAIT
Man Cal
7pH
60 s.
Error
7 pH
100% ok
7.00 pH
Man Cal
7.00 pH
4 pH
100% ok
to exit the calibration (after 4 seconds the pump will
to exit the calibration (after 4 seconds the pump will
to insert the value of my solution, then press
EN
Operation
WAIT
60 s.
Error
4.00 pH
to set
and the display
and the display
to confirm
12